6.7.2. Analog Output Step Test

This test can be used to check the accuracy and proper operation of the analog outputs.
The test forces all four analog output channels to produce signals ranging from 0% to 100%
of the full scale range in 20% increments. This test is useful to verify the operation of the
data logging/recording devices attached to the analyzer.
To begin the Analog Output Step Test press:
SAMPLE
RANGE = 500.0 PPB
< TST TST > CAL
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK
SETUP X.X
SECONDARY SETUP MENU
COMM VARS
DIAG
SETUP X.X
ENTER DIAG PASS: 818
8
1
8
044100102 Rev A
RANGE = 500.0 PPB
NOX=X.X
SETUP
PRIMARY SETUP MENU
MORE
EXIT
SECONDARY SETUP MENU
DIAG
EXIT
ENTER DIAG PASS: 818
8
ENTR
EXIT
NOX=X.X
SETUP
MORE
EXIT
EXIT
ENTR
EXIT
DIAG
SIGNAL I / O
PREV NEXT
JUMP
DIAG I / O
Test Signals Displayed Here
PREV NEXT
JUMP
EXAMPLE
DIAG I / O
JUMP TO: 5
0
5
DIAG I / O
CAL_LED = ON
PREV NEXT JUMP
Pressing the PRNT key will send a formatted printout to the serial port and can be
captured with a computer or other output device.
DIAG
SIGNAL I / O
NEXT
DIAG
ANALOG OUTPUT
PREV
NEXT
DIAG AOUT
ANALOG OUTPUT
0%
DIAG AOUT
ANALOG OUTPUT
[0%]
Pressing the key under "0%" while performing the test will
pause the test at that level. Brackets will appear around
the value: example: [20%] Pressing the same key again
will resume the test.
Operating Instructions
Use the NEXT & PREV
keys to move between
ENTR
EXIT
signal types.
Use the JUMP key to
go directly to a
specific signal
See Appendix A-4 for
PRNT EXIT
a complete list of
available SIGNALS
Enter 05 to Jump
(CAL_LED)
ENTR
EXIT
Exit to return
to the
ON
PRNT EXIT
DIAG menu
ENTR
EXIT
ENTR
EXIT
analog output
EXIT
0% - 100%
returns to the
DIAG menu
EXIT
to Signal 5:
Performs
step test.
Exit-Exit
59

Hide quick links:

Advertisement

Table of Contents

Troubleshooting

loading

Table of Contents