Table 8-1: Daily Activity Matrix - Teledyne 400E Instruction Manual

Ozone analyzer
Table of Contents

Advertisement

EPA Protocol Calibration
In addition, an independent precision check between 0.08 and 0.10 ppm may be
required at least once every two weeks. Table 8-3 summarizes the quality
assurance activities for routine operations. A discussion of each activity appears in
the following sections.
To provide for documentation and accountability of activities, a checklist should be
compiled and then filled out by the field operator as each activity is completed.
Characteristic
Shelter Temperature
Sample Introduction
System
Recorder
Analyzer Operational
Settings
Analyzer Operational
Check
Precision Check
200

Table 8-1: Daily Activity Matrix

Acceptance Limits
Mean temperature
between 22°C and 28°C
(72°F and 82°F), daily
fluctuations not greater
than ± 2°C.
No moisture, foreign
material, leaks,
obstructions; sample line
connected to manifold.
Adequate ink supply and
chart paper.
Legible ink traces.
Correct settings of chart
speed and range
switches.
Correct time.
Flow and regulator
indicators at proper
settings.
Temperate indicators
cycling or at proper
levels.
Analyzer in sample
mode.
Zero/span controls
locked.
Zero and span within
tolerance limits as
described in Subsec.
9.1.3 of Sec. 2.0.9
(Ref. 11).
Assess precision as
described in Sec. 2.0.8
(Ref. 11).
Model 400E Ozone Analyzer
Frequency and
Method of
Measurement
Check thermograph
chart daily for
variations not greater
than ± 2°C (4°F).
Weekly visual
inspection.
Weekly visual
inspection.
Weekly visual
inspection.
Level 1 zero and span
every 2 weeks; Level
2 between Level 1
checks at frequency
desired by user.
Every 2 weeks,
Sec. 2.0.8 (Ref. 11).
Instruction Manual
Action if
Requirements are
not Met
Mark strip chart for the
affected time period.
Repair/adjust temp
control.
Clean, repair or replace
as needed.
Replenish and chart
paper supply
Adjust recorder time to
agree with clock note
on chart.
Adjust or repair as
needed.
Isolate source error,
and repair.
After corrective action
recalibrate analyzer.
Calculate, report
precision,
Sec. 2.0.8 (Ref. 11).
P/N 04316 Rev: B