Calibrator Accuracy Test

Measure the accuracy of the Calibrator output.
Test Limits
Voltage ± (1.5 mV + voltage*0.005)
Table 4-5. Calibrator Output Test Limits
Equipment List
• 11408C, 50 ohm BNC feed thru
• 34401A, Digital Multi Meter
• Dual Banana Plug
Manual Procedure
1 Connect the Calibrator output of the DCA to the 11408C.
2 Connect the other end of the 11408C to the input of the 34401A with a BNC to dual banana plug.
3 Push the default setup button on the DCA.
4 Using the user interface on the DCA select Calibrate, Front Panel Cal output
5 Set the calibrator output to each of the first data points.
Input
Voltage
Min Test Limits
- 2 VDC
-2.0115 VDC
- 1.667 VDC
-1.676835 VDC
- 1.333 VDC
-1.341165 VDC
- 1 VDC
-1.0065 VDC
- 0.667 VDC
-0.671835 VDC
- 0.333 VDC
-0.336165 VDC
0 VDC
-0.0015 VDC
+ 0.333 VDC
0.329835 VDC
+ 0.667 VDC
0.662165 VDC
+ 1 VDC
0.9935 VDC
+ 1.333 VDC
1.324835 VDC
+ 1.667 VDC
1.657165 VDC
+ 2 VDC
1.9885 VDC
Max Test Limits
-1.9885 VDC
-1.65717 VDC
-1.32484 VDC
-0.9935 VDC
-0.66217 VDC
-0.329835 VDC
0.0015 VDC
0.336165 VDC
0.967335 VDC
1.0065 VDC
1.341165 VDC
1.676835 VDC
2.0115 VDC
Performance Verification

Calibrator Accuracy Test

4-29