Agilent Technologies Infinium DCA 86100A Service Manual page 25

Wide-bandwidth oscilloscope
Table of Contents

Advertisement

N O T E
If you can't make your license file available to the 86100C, you can manually enter a 60 digit key
code. Refer to the steps below.
Enter the license key
Although it is faster to use the file, manually entering the key code provides an alternative that
manually (optional)
fully fulfills the installation requirements.
1 Obtain the 60 digit key code by using a text editor, such as Notepad, and view the contents of the
license file. The license filename ends in the extension ".lic"
Here is an example of the contents of a license file:
2 After opening the file, look for the 60-digit key code after the header "SIGN=" and in between the
quotation marks e.g.:
00A2 E3A3 436C B985 ACE6 7732 036F AF00 9208 23B4 AEEC 1660 F8D9 EBD1 219A
N O T E
There may be more that one 60-digit code because of multiple licenses. Be sure to enter all sets
of 60-digit codes.
3 After finding the key code, click the Install New License dialog box to enter your code and
activate your feature.
N O T E
Because the letters in the key code are not case sensitive, you can enter letters in either upper or
lower case. The key code is in hexadecimal using digits 0 through 9 and letters a through f.
Transferring a License Transferring a license from one DCA to another is not permitted. A separate license must be pur-
chased for each DCA that requires that feature.
Delete Licenses From
Deleting licenses is required when refurbishing a DCA for resale. To remove all installed feature
the DCA
licenses, follow the steps below:
1 Connect a keyboard to the instrument.
2 On the DCA click Exit on the File menu to close the 86100C application.
3 When asked "are you sure you want to exit 86100C", click Yes.
4 Click "Start" and select "Run..."
5 Type the following in the Run window:
6 Click OK to run the batch file.
7 At the on-screen message referring to license removal, follow the instructions to proceed.
8 Press any key such as the Enter key.
9 To confirm that the licenses are removed, start the DCA application by double clicking the
86100C infiniium DCA shortcut on the DCA desktop.
10 Touch/click Utilities, Licenses...
11 Check under the "Features That Require a License" box and confirm that the status for each
feature listed is "Not Installed".
FEATURE AdvancedWaveformAnalysis agilent 1.0 permanent uncounted
HOSTID=003064023c0a TS_OK SIGN="00A2 E3A3 436C B985 ACE6 7732 036F AF00 9208
23B4 AEEC 1660 F8D9 EBD1 219A"
c:\windows\i386\$OEM$\RemoveLicenses.bat
Warning!
You are about to remove a license.
Press any key to continue, or ctrl-c to break out...