Chapter 3: Testing Performance

To test the dc calibrator

To test the dc calibrator

The Aux Out BNC on the front panel is used for self-calibration and probe calibration. Though
calibrator accuracy is not specified in the performance specifications, it must be within limits in
order to provide accurate self-calibration.
Test Limits: -2.5 v to +2.5 v, Accuracy ±0.2% of delta voltage output
Equipment Required
Equipment
Digital Multimeter
Cable
Adapter

Procedure

Connect the multimeter to the front panel Aux Out BNC.
1
Use the BNC cable and the BNC to banana plug adapter.
Enable the graphical interface.
2
Use the mouse to click on the button in the upper right-hand corner of the display. See figure 3-1.
Figure 3-1
Enabling the graphical interface
Press Default Setup to set the oscilloscope to default conditions.
3
Select Calibration from the Utilities menu.
4
Select DC from the Aux Output drop-down list box.
5
See figure 3-2.
3–4
Critical Specifications
0.1 mV resolution, better than 0.1 % accuracy
BNC
BNC (f) to banana (m)
Click the mouse with the
pointer on the right-hand
button to enable the
graphical interface
Recommended Model/Part
Agilent 34401A
Agilent 10503A
Agilent 1251-2277