To Replace the A3 Backlight Inverter

C A U T I O N
Electrostatic discharge (ESD) can damage or destroy electrostatic components. All work on
electronic assemblies should be performed at a static-safe work station. See
Discharge Information" on page 1-8
1 Disconnect the power cord from the instrument.
WA R N I N G
Opening covers or removing parts is likely to expose dangerous voltages. Disconnect the
instrument from all voltages before it is opened.
2 Remove the trim strips from both sides of the front panel, as shown in
Figure 7-27. Removing the Trim Strips
3 Use the T-15 TORX driver to remove the four screws (two on each side) that secure the front panel
to the mainframe, as shown in
for more information on preventing ESD.
Figure
7-28.
Replacing Instrument Assemblies

To Replace the A3 Backlight Inverter

"Electrostatic
Figure
7-27.
7-21