3 Set the 83712B to 2.5 GHz, +2.50 dBm and turn the RF output on.
4 Set the DCA for the following.
5 Set the RF switching for DCA triggers Off and RF output to DCA channel 2.
6 Check to see if the DCA is triggering with the TER command.
7 Set the RF switching for RF output to DCA Module Triggers, Set the Trigger source of the DCA
8 Set the RF switching for RF output to DCA Module Triggers, Set the Trigger source of the DCA
9 Set the RF switching for RF output to DCA Front Panel Trigger output and set the Trigger source
10 If all steps are a pass then the whole test is a pass. Otherwise the test is a failure.
Data Storage
• Front Panel Trigger
• Left Module Trigger
• Right Module Trigger
• Freerun mode
Performance Verification
Trigger Port Check
a "DUT Front Trigger" to the Front panel trigger of the 86100A/B/C.
b "DUT lmod Trigger" to the Left Module trigger of the DUT.
c "DUT rmod Trigger" to the Right Module trigger of the DUT.
d "DUT Ch 2" to the Channel 2 input of the DUT.
a Default Setup.
b Turn off all channels except for channel 2.
c 50 mV/div on Channel 2.
d 100 ps/div.
e Trigger source to freerun.
to the Right Module. Execute
to the Left Module. Execute
of the DCA to front panel. Execute
4-14
Step 6
again.
Step 6
again.
Step 6
again.