Performance Verification
Calibrator Accuracy Test
a On the 34401A read the DC voltage and record the value. Compare this value with the speci-
b Repeat
Software
1 Prompt user to connect the cable labeled "DUT cal output" to the Calibrator output of the DUT.
2 Execute a default setup. Set the calibrator output to the desired measured parameter listed above
in the specifications.
3 Take a DC voltage reading on the 34401A
4 Repeat
5 Compare the measured and calculated values with the specifications listed above to determine
Pass or Fail.
6 Data should be presented in a graphical format with data point on the x-axis and the DMM reading
on the y-axis. The calculated parameters should be presented in a tabular format either on the
graph or as a separate sheet.
Data Storage
• Parameter setting
• Measured Value
• Min Spec
• Max Spec
• Calculated Value
• Min Spec
• Max Spec
4-30
fications listed above.
Step 5
for all of the data points.
Step 2
until all the data points are completed.