Replacing Instrument Assemblies
To Replace the A9 Hard Drive
13 Connect a keyboard to the instrument.
14 Turn the instrument on.
15 Repeatedly press the function key F8 as it boots up.
16 If done correctly the Windows Advanced Options Menu will eventually be displayed as shown
below:
Windows Advanced Options Menu
Please select an option
Use the up and down arrow keys to move the highlight your choice.
17 Select "Return to OS Choices Menu" and hit Enter.
N O T E
If instead you see the "Welcome to the Windows XP Setup Wizard", click Next and do not accept
the agreement. Click Next, Exit the Setup, Close the Fatal Error dialog box and start the process
over by going to
18 The following text will be displayed:
19 Press the down arrow key on the keyboard to select Emergency Backup And Recovery. Then,
press the Enter key. "The FreeDOS Project" copyright message and the prompt: "Press Any Key
to Continue..." will be displayed.
20 Pressing any key will display the following text:
7-40
Safe Mode
Safe Mode with Networking
Safe Mode with Command Prompt
Enable Boot Logging
Enable VGA Mode
Last Known Good Configuration (your most recent settings that worked)
Directory Services Restore Mode (Windows domain controllers only)
Debugging Mode
Disable automatic restart on system failure
Start Windows Normally
Reboot
Return to OS Choices Menu
Step 13
and repeating procedure.