Performance Verification

Front Panel Check

Front Panel Check

Using the built in self tests, ensure the front panel is working properly by exercising all buttons,
knobs, and LEDs.
Manual Procedure
1 Touch/click Help, About 86100..., and press Local 5 times. This allows access to the service
mode.
2 Touch/click Utilities, Service, Self Test.
3 Chose Front Panel and click Start Test.
4 The screen will display a layout of the mainframe front panel showing all the buttons and knobs
in red except for the power button. The 86100A or B do not have a Jitter Button.
5 Verify that all keys toggle and that all knobs rotate left and right. Make sure all red buttons become
green. The color of the knobs will turn yellow when you rotate left or right then green when you
rotate the other direction.
6 Press All On for the LED Test. Make sure that all the LED light up properly with correct color.
7 Press All Off for the LED Test. Make sure that all LED do not light up except the power button.
8 Press Exit Front Panel Test to exit.
Software
Prompt user to execute the manual procedure. Query the user for when this step is completed
(yes/no). Default should be no.
Data Storage
Store Pass and Fail based on the user input. Store the date and time the test was completed.