Agilent Technologies Infinium DCA 86100A Service Manual page 87

Wide-bandwidth oscilloscope
Table of Contents

Advertisement

Table 4-10. 86100C Test Conditions and Data Points for Option 001 Trigger Jitter Test
Frequenc
y
Delay
3.0 GHz
24 ns
5.0 GHz
24 ns
8.0 GHz
24 ns
10 GHz
24 ns
12 GHz
24 ns
12 GHz
100 ns
13 GHz
24 ns
14 GHz
24 ns
15 GHz
24 ns
3.0 GHz
41.1
ns
11 For each data point do the following:
12 Set the DCA to 100 mV/div, (0.5 / frequency) ps/div and delay listed in
depending on the model number under test.
13 Set the 83712B to the new frequency.
14 Adjust the 83712B until required input voltage is read on channel 2. Use a binary search to achieve
this result to within ±2 mV.
15 Set the trigger mode as required.
16 Use a marker to find the delay of the 2nd rising edge at 0 V currently on the screen.
17 Set the DCA to listed vertical and horizontal scale settings.
18 Set the DCA to the delay setting from
19 Setup the Histogram window to measure the RMS jitter in the center of the screen. Refer to the
histogram window information above.
20 Push the stop button, then the clear screen button, followed by the run button.
21 Let the DCA collect histogram data until there are 1000 or more hits.
22 After the 1000 hits have been collected then read the histograms standard deviation (this is the
RMS jitter reading).
23 Repeat
Step 11
for each parameter in the specifications above.
24 Compare each parameter with the specifications above.
Input
Horizon
Vertical
Voltage
tal Scale
Scale
(Vpp)
20 ps/
5 mV/div
192 mV
div
5 ps/div
5 mV/div
189 mV
5 ps/div
5 mV/div
180 mV
2 ps/div
10 mV/div
356 mV
2 ps/div
10 mV/div
348 mV
10 ps/
50 mV/div
348 mV
div
2 ps/div
10 mV/div
340 mV
2 ps/div
12.5 mV/
420 mV
div
2 ps/div
12.5 mV/
415 mV
div
20 ps/
5 mV/div
192 mV
div
Step
16.
Performance Verification
Trigger Jitter Test, Option 001
Trigger
Histogra
Histogra
Mode
m Height
m Width
2-12 GHz
± 300 uV
± 100 ps
2-12 GHz
± 300 uV
± 25 ps
2-12 GHz
± 300 uV
± 25 ps
2-12 GHz
± 600 uV
± 10 ps
2-12 GHz
± 600 uV
± 10 ps
2-12 GHz
± 600 uV
± 20 ps
2-12 GHz
± 600 uV
± 10 ps
2-12 GHz
± 800 uV
± 10 ps
2-12 GHz
± 800 uV
± 10 ps
Pattern
± 300 uV
± 100 ps
Trigger
Table 4-9
or
Table 4-10
4-37

Advertisement

Table of Contents
loading

This manual is also suitable for:

Infinium dca 86100bInfinium dca 86100c

Table of Contents