Instrument Modifications
Replacing A7A2, A8, or A9 Assemblies
Copy re-direct files to DCA hard drive
42 Download Redirect.exe from the support web site to a temporary memory location such as a USB
memory stick.
43 Double click on Redirect.exe. This is a self extracting archive; by executing the file, it will
automatically copy the redirect files listed below to C:\scope\touchscreen on the DCA. During the
file copy process, it will overwrite the existing tbmorph.exe file.
• Microsoft.VC80.CRT.manifest
• msvcm80.dll
• msvcp80.dll
• msvcr80.dll
• tbmorph.exe
44 Reboot the DCA by clicking on Start, Turn Off Computer, Restart.
Touch Screen Calibration:
45 After the DCA is re-booted, click on Utilities, Touch Screen Config...
46 You should see the screen as shown in
in
Figure
6-10, then you may have to repeat the process to reinstall the new drives and redirect
files.
Figure 6-9. Touch Screen Driver for the 0960-2590 TS Controller
6-10
Figure
6-9. If you see the old configuration screen as shown