Figure 7-5 Test connection from calibrator to instrument and digital multimeter

4 Adjust the output of the calibrator and observe the reading display on
5 Wait a few seconds for the measurement to settle. Compare the
6 Continue to check voltage measurement accuracy with the remaining
7 When you have finished checking all of the power supply setting values,
8 Repeat the same verification procedures for channel 2.
162
the multimeter.
measurement results displayed on the multimeter to the corresponding
test limits shown in Table 7- 3.
Volts/div setting values in Table 7- 3.
disconnect the power supply from the oscilloscope.
Agilent U1602A/U1604A User's and Service Guide