Instrument Refurbishment
Refurbishing an 86100C (Alternate Method)
N O T E
F is not a choice that is displayed on the screen, it is a hidden function used only in the factory and
field to create a new backup image of the C partition.
N O T E
The system will perform a quick integrity check of the file structure on the C: partition. It will then
copy the C: partition to an image file and store it on the System Recovery partition. This process
will overwrite the original Factory Recovery Image.
46 Press C to continue or E to exit. Read the on-screen message then press C to continue or E to exit
one more time.
47 Wait for the factory image backup process to complete.
48 Even though the on-screen instructions say to press CLT+ALT+DEL, don't do it. Power off the
DUT using power button.
N O T E
DO NOT press CLT+ALT+DEL to restart the DCA.
C A U T I O N
Do not attempt to power up the DCA again. Doing so would nullify the Re-Seal OS process which
prepares the OS for the first customer run experience and potentially violate the license agreement
with Microsoft. The new owner should power up the DCA for the first time so that they can agree
to the licensing requirements of Windows XP.
49 Cancel License Entry at Agilent Software Licensing.
page
N O T E
If the refurbished DCA held software licenses before the hard drive was replaced, then the entries
for these software licenses need to be cancelled at the Agilent Software Licensing web site.
5-12
1-20.
Refer to "Cancel ASL License Entry" on