Performance Verification

Module Calibration Check

Module Calibration Check

Calibrate a known good module in the left slot and right slot. This ensures that there are no prob-
lems in the mainframe preventing the module calibration from passing.
Equipment List
• Two 83483A Dual Electrical Module
Manual Procedure
1 Insert a module into the left or right slot.
2 Touch/click Calibrations drop-down menu.
3 Select All Calibrations.
4 Select the Modules tab and Calibrate the left or right module as appropriate.
5 Wait for the calibration to complete.
6 Repeat the procedure until both slots have been checked and that module calibration passed.
4-20