Instrument Modifications

A13 Acquisition Board PLD Header Modification (86100A Only)

6 Secure the A16 LS-120 Board to the A10 Floppy Drive with two, T-10 Torx Screws. Refer to
Figure
7 Secure the ribbon cable W9 to the A16 LS-120 Board. Refer to
N O T E
Ensure that the red stripe on the ribbon cable, W9 is towards the fan assembly when connection
the cable to LS-120.
8 Reassemble the Mainframe Cover.
9 Perform a Read/Write test of the Floppy Drive.
• Power on the instrument, and insert any 3.5 floppy disk into the floppy drive.
• At the touchscreen select the File pull-down menu, and select "Save, Instrument Setup."
• Select "Look In" 3.5 Floppy (A:), touch/click Save.
• Select "Open Instrument Setup" and recall the file that was just saved.
• The saved setup should be recovered properly.
A13 Acquisition Board PLD Header Modification
(86100A Only)
1 Remove the instrument cover.
2 Remove the A13 Acquisition Board.
Counter Assembly (Option 001)" on page
3 Place the jumper, shown in
5, shown in
Figure 6-20. Jumper, E2660-01201
6-20
6-18.
Refer to "To Remove the Instrument Cover" on page
Figure
Figure
6-21. The jumper key should face towards the front panel.
Refer to "To Replace the A13 Acquisition and 86100C A14
7-45.
6-20, on P12 so that the resistors are connected to pins 1, 4, and
Figure
6-18.
7-3.