Table 4 Equipment Required To Test Digital Channel Threshold Accuracy - Agilent Technologies 6000 Series Service Manual

Hide thumbs Also See for 6000 Series:
Table of Contents

Advertisement

6000 Series Oscilloscopes Service Guide
Table 4
Equipment Required to Test Digital Channel Threshold
Accuracy
Equipment
Digital Multimeter
Oscilloscope Calibrator
BNC-Banana Cable
BNC Tee
BNC Cable
BNC Test Connector,
8-by-2
Probe Cable
1 Turn on the test equipment and the oscilloscope. Let them
warm up for 30 minutes before starting the test.
2 Set up the oscilloscope calibrator.
a Set the oscilloscope calibrator to provide a DC offset
voltage at the Channel 1 output.
b Use the multimeter to monitor the oscilloscope calibrator
DC output voltage.
3 Use the 8-by-2 test connector and the BNC cable assembly to
connect digital channels D0-D7 to one side of the BNC Tee.
Then connect the D0-D7 ground lead to the ground side of
the 8-by-2 connector. See
Testing Performance
Critical
Recommended Model/Part
Specifications
0.1 mV resolution,
Agilent 34401A
0.005% accuracy
DC offset voltage
Fluke 5820A
6.3 V
Agilent 11001-60001
Agilent 1250-0781
Fluke 50
User-built (See
Agilent 01650-61607
Figure
2.
2
cable, P/N 686318
page
29.)
33

Advertisement

Table of Contents
loading

Table of Contents