Agilent Technologies Infinium DCA 86100A Service Manual page 82

Wide-bandwidth oscilloscope
Table of Contents

Advertisement

Performance Verification
Trigger Jitter Test, Standard
Equipment List
• 83712B 20 GHz Synthesized CW generator opt 1E1, 1E5, 1E8, 1E9
• 11636B Power Splitter
• RF cables and adaptors as necessary
• DCA with an 83483A module in the left slot
Manual Procedure
Connection Setup
1 Connect the 11636B to the RF output of the 83712B.
2 Connect one output of the 11636B to the front panel trigger input of the DCA.
3 Connect the other output of the 11636B to the channel 2 input of the DCA.
DCA Setup
4 Push the default setup button.
5 Turn off all channels except for channel 2.
6 Set the Horizontal Scale to 200 ps/div.
7 Set the trigger source to the front.
Table 4-7. Test Conditions and Data Points for Standard Trigger Jitter Test
Take Measurements
8 For each data point, do the following:
9 Set the DCA to 100 mV/div, (0.5 / frequency) ps/div and delay listed in
10 Set the 83712B to the new frequency.
11 Adjust the 83712B until required input voltage is read on channel 2. Use a binary search to achieve
this result to within ± 2 mV.
12 Set the trigger mode as required.
13 Use a marker to find the delay of the 2nd rising edge at 0 V currently on the screen.
14 Set the DCA to listed vertical and horizontal scale settings.
15 Set the DCA to the delay setting from
16 Setup the Horizontal Histogram window to measure the RMS jitter in the center of the screen.
4-32
Dela
Horizonta
Freq
y
l Scale
2.5 GHz
24 ns
20 ps/div
2.5 GHz
200
20 ps/div
ns
2.75
24 ns
20 ps/div
GHz
3.2 GHz
24 ns
20 ps/div
Input
Vertical
Voltage
Trigger
Scale
(Vpp)
Sense
5 mV/div
193 mV
Normal
5 mV/div
193 mV
Normal
5 mV/div
193 mV
High
5 mV/div
191 mV
Normal
Step
13.
Histogram
Histogra
Height
m Width
± 300 uV
± 100 ps
± 300 uV
± 100 ps
± 300 uV
± 100 ps
±300 uV
± 100 ps
Table
4-7.

Advertisement

Table of Contents
loading

This manual is also suitable for:

Infinium dca 86100bInfinium dca 86100c

Table of Contents