Internal Temperature Check - Agilent Technologies Infinium DCA 86100A Service Manual

Wide-bandwidth oscilloscope
Table of Contents

Advertisement

Manual Procedure
1 Using the touchscreen click on the Help drop down menu then select About 86100...
2 In the window that pops up look at the internal temperature inside the system configuration.
3 Compare the temperature of the DCA to the test limits of 20 to 40 degrees C.
Software
Data Storage
Performance Verification

Internal Temperature Check

Internal Temperature Check

The purpose of this test is to verify that the internal temperature of the DCA is within the test
limits before performance verification data is collected.
Query the temperature of the DCA and compare temperature to the test limits.
Store the temperature of the DCA and identify it with the date and time the test was run.
4-12