Manual Procedure
1 Using the touchscreen click on the Help drop down menu then select About 86100...
2 In the window that pops up look at the internal temperature inside the system configuration.
3 Compare the temperature of the DCA to the test limits of 20 to 40 degrees C.
Software
Data Storage
Performance Verification
Internal Temperature Check
Internal Temperature Check
The purpose of this test is to verify that the internal temperature of the DCA is within the test
limits before performance verification data is collected.
Query the temperature of the DCA and compare temperature to the test limits.
Store the temperature of the DCA and identify it with the date and time the test was run.