Agilent Technologies Infinium DCA 86100A Service Manual page 121

Wide-bandwidth oscilloscope
Table of Contents

Advertisement

Figure 6-10. Touch Screen Driver for the older 1150-7826 TS Controller
47 Perform a Touch Screen Calibration.
48 Refer to Service Note 86100C-06 for touch screen support information.
49 Instrument functionality testing should be performed after installing this kit. Refer to the 86100A/
B/C Service Guide, part number 86100-90075 for more information.
N O T E
After the completion of the touch screen controller retrofit, the A9 hard drive must have a
recovery image created on the hidden partition. This is because of the new software drivers that
were installed.
Create New Backup
Creates a backup copy of the C partition with all the recently created files, and stores it on a hid-
Image of Drive C
den partition.
1 Connect a keyboard to the instrument.
2 Turn the instrument on.
3 Repeatedly press the function key F8 as it boots up.
4 If done correctly the Windows Advanced Options Menu will eventually be displayed as shown
below:
Windows Advanced Options Menu
Please select an option
Safe Mode
Safe Mode with Networking
Safe Mode with Command Prompt
Enable Boot Logging
Enable VGA Mode
Last Known Good Configuration (your most recent settings that worked)
Directory Services Restore Mode (Windows domain controllers only)
Debugging Mode
Instrument Modifications
Replacing A7A2, A8, or A9 Assemblies
6-11