Agilent Technologies Infinium DCA 86100A Service Manual page 107

Wide-bandwidth oscilloscope
Table of Contents

Advertisement

Windows Advanced Options Menu
Please select an option
Use the up and down arrow keys to move the highlight your choice.
41 Select "Return to OS Choices Menu" and hit Enter.
N O T E
If instead you see the "Welcome to the Windows XP Setup Wizard", click Next and do not accept
the agreement. Click Next, Exit the Setup, Close the Fatal Error dialog box and start the process
over by going to
42 The following text will be displayed:
43 Press the down arrow key on the keyboard to select Emergency Backup And Recovery. Then,
press the Enter key. "The FreeDOS Project" copyright message and the prompt: "Press Any Key
to Continue..." will be displayed.
44 Pressing any key will display the following text:
45 Press F at this point to create a factory backup image file.
Safe Mode
Safe Mode with Networking
Safe Mode with Command Prompt
Enable Boot Logging
Enable VGA Mode
Last Known Good Configuration (your most recent settings that worked)
Directory Services Restore Mode (Windows domain controllers only)
Debugging Mode
Disable automatic restart on system failure
Start Windows Normally
Reboot
Return to OS Choices Menu
Step 30
and repeating procedure.
Instrument Refurbishment
Refurbishing an 86100C (Alternate Method)
5-11