Replacing Instrument Assemblies
To Replace the A2 Flat Panel Display
To Replace the A2 Flat Panel Display
1 Follow the instructions for removing the front panel on page 7-6.
2 Use a T-10 TORX driver to remove the four screws that secure the display to the front panel, as
shown in
Figure
7-25.
Figure 7-25. Removing the Display Screws
3 Carefully lift the display out of the front panel.
Figure 7-26. Removing the Display
4 To re-install the display, perform the above steps in the reverse order.