Linearity/Attenuation Accuracy Test; Figure D-6 Total Insertion Loss Test Setup 2, Option 350 - Agilent Technologies 8156A Operating And Programming Manual

Attentuator
Table of Contents

Advertisement

Figure D-6
Total Insertion Loss Test Setup 2, Option 350
Specifications
Linearity
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Performance Tests
Performance Test
7. Enable the attenuator output and record the power meter reading
(in dB) in the Test Record and check that it is within
specifications.
II. Linearity/Attenuation Accuracy Test
Agilent 8156A
Option 100
Option 101
Option 121
Option 201
Option 221
Option 350
Carry out the following Attenuation Accuracy tests at 1310nm and
1550nm with single-mode fibers using the equipment listed
previously.
182
< ± 0.2dB
< ± 0.1dB
< ± 0.1dB
< ± 0.1dB
< ± 0.1dB
< ± 0.1dB