Agilent Technologies Infinium DCA 86100A Service Manual page 165

Wide-bandwidth oscilloscope
Table of Contents

Advertisement

7-47.
Figure 7-47. Press Delete to start CMOS Setup Utility
28 Use the arrow keys to select Load 86100C Setting as shown in
Setting is you are setting up an 86100B instrument. Press the Enter key and then the Y key to begin
the load process.
29 When the load process completes, press F10. When the message shown in
press the Y key followed by the Enter key. This exits the CMOS Setup Utility.
30 Reboot the instrument to complete the BIOS flash process.
Figure 7-48. Loading the 86100C Settings
Replacing Instrument Assemblies
To install the motherboard
Figure
7-48. Select Load 86100B
Figure 7-49
appears,
7-33