Touchscreen Calibration
Calibrate the touchscreen by following the on-screen instructions. The procedures differ some-
what from one DCA model number to another.
Equipment List
Agilent Part Number 1535-5214, Touch screen stylus
Manual Procedure
With the DCA powered up touch/click as described for each model number:
86100A:
1 Click on Utilities, Touch Screen Config..., Calibration..., Calibrate
2 Using a stylus pen, follow the on-screen instructions of touching each of the three crosses on the
screen
86100B:
1 Click on Utilities, Touch Screen Calibration...,
2 Using a stylus pen, touch the center of each cross as it appears. Calibration will terminate if no
touch is received within 20 second
86100C:
1 Click on Utilities, Touch Screen Config...,
2 On the Pointer Device Properties window click on the Calibrate button to start the calibration
process
3 Use stylus pen to touch the center of each cross as it appears. Calibration will terminate if no touch
is received within 10 seconds
4 After completing the calibration, tap the OK button to complete and close the Pointer Device
Properties window
Adjustments
Touchscreen Calibration
3-9