To Replace The A4 Pc Motherboard - Agilent Technologies Infinium DCA 86100A Service Manual

Wide-bandwidth oscilloscope
Table of Contents

Advertisement

Replacing Instrument Assemblies

To Replace the A4 PC Motherboard

To Replace the A4 PC Motherboard

Photographs of the 86100A are used in the following steps. Although some of the components
are different in the 86100B/C, the removal procedure is virtually identical. Removing the A4
assembly in the 86100C does not require the removal of the fan.
When replacing the A4 assembly with a new version, be sure to follow the instructions at the end
of this replacement procedure.
Replacement of the A4 PC Motherboard will require that any installed software license be
"keyed" to the new motherboard. Option 200, Enhanced Jitter Software, is an example of a soft-
ware option. The motherboard contains a unique Host ID number which is used in the license
key generation process.
instructions on regenerating a software license file after a motherboard replacement.
1 Remove the instrument's cover.
2 Remove four T-10 TORX screws that secure the fan, as shown in
of the mainframe.
Figure 7-32. Removing the Fan (86100A/B Only)
3 Remove two T-10 TORX screws that secure the bar to the mainframe, as shown in
7-24
Refer to "To install the motherboard" on page
Refer to "Software License Maintenance, 86100C" on page 1-16
Refer to "To Remove the Instrument Cover" on page
7-31.
for
7-3.
Figure
7-32, then lift the fan out
Figure
7-33.

Advertisement