Replacing Instrument Assemblies
To Replace the A13 Acquisition and 86100C A14 Counter Assembly (Option 001)
N O T E
On the 86100C the switch is located closer to the Acquisition assembly; however, the same cables
should be removed.
3 Turn the instrument on its side so that you can access the screws that secure the A13 Acquisition
board, then remove the two screws, as shown in
Figure 7-62. Removing the A13 Acquisition Board Screws
4 Slide the board to the right to free it from the locking posts.
Figure 7-63. Sliding the Board to the Right
5 Pull the board out a few inches, then disconnect the W24 ribbon cable and the cable from the SW2
Transfer Switch, as shown in
7-46
Figure
7-62.
Figure 7-64
and
Figure
7-65.