To Replace the A7A2 Touch Screen

Before replacing the A7A2 assembly, refer to
on page
6-3.
1 Follow the instructions for removing the front panel on page 7-6, and the display on page 7-20.
2 Using a T-10 TORX driver, remove two screws that secure the touch screen board bracket, as
shown in
Figure
7-53.
Figure 7-53. Removing the Touch Screen Screws
3 Lift the touch screen out of the display assembly, as shown in
Figure 7-54. Lifting out the Touch Screen
4 To re-install the touch screen, perform the above steps in the reverse order.
Replacing Instrument Assemblies
To Replace the A7A2 Touch Screen
refer to "Replacing A7A2, A8, or A9 Assemblies"
Figure
7-54.
7-37