To Replace the A7 Front-Panel Keyboard

1 Follow the instructions for removing the front panel on page 7-6, and the display on page 7-20.
2 Turn the front panel face up.
3 Pull off all knobs, as shown in
Figure 7-50. Removing the Knobs
4 Turn the panel face down.
5 Use a T-10 TORX driver to remove the two screws from the touch screen board, as shown in
Figure
7-51, then lift bracket out of front panel.
Figure 7-51. Removing the Touch Screen Screws
6 Remove the eight screws that secure the keyboard to the front panel, as shown in
Figure
7-50.
Replacing Instrument Assemblies
To Replace the A7 Front-Panel Keyboard
Figure
7-52.
7-35