Replace Parts of the Inline Filter

Tools required
p/n
8710-0510
Parts required
p/n
5023-0271
Stuck Capillary in Multi Purpose Valve
C AU T I O N
Shortcutting the inline filter by directly connecting its right capillary to valve port 5 can
damage the Multi Purpose Valve.
The size/position of this capillary in its fitting is incompatible, so it may get stuck
irreversibly to the valve.
The inline filter can be cleaned using the back-flush function in the user interface of your
N O T E
Agilent instrument control software.
Only use the back-flush option, if an inline filter is installed.
1260 Infinity II Flexible Pump User Manual
Description
Wrench open 1/4 — 5/16 inch
Description
Frit 0.3 µm for inline filter, 5/pk
Do not shortcut the filter by directly connecting its right capillary to valve port 5 in
case the inline filter cannot or shall not be used.
Use Capillary ST 0.17 x 120 mm, SLV/SV (5067-5416) instead.
Maintenance

Replace Parts of the Inline Filter

7
173