  • Page 1 Agilent 1260 Infinity II High Temperature Sample Preparation System User Manual Agilent Technologies...
  • Page 2 Notices Warranty © Agilent Technologies, Inc. 2016 Trademarks No part of this manual may be reproduced in The material contained in this document is Windows and MS Windows are U.S. any form or by any means (including provided “as is,” and is subject to being...
  • Page 3: Health And Safety

    Health and Safety Agilent 1260 Infinity II High Temperature Sample Preparation System The use of this instrument does not entail any hazard if used in accordance with the instructions stated within the Operator’s manual. The protection provided by this equipment may be impaired if it is used with accessories not provided or recommended by Agilent or used in a manner not specified in this manual.
  • Page 4: Table Of Contents

    Contents NOTICES HEALTH AND SAFETY GENERAL INFORMATION Overview Specifications Unpacking Location Utility Requirements Safety Precautions SYSTEM OVERVIEW INSTALLING THE MODULE Preparing the system SYSTEM OPERATION Power On Shaker Temperature Controller Safety Over-Temperature Cut-Out Hand Pipettor SAMPLE PREPARATION Procedure Cleaning the Frits CLEANING AND DECONTAMINATION PROCEDURE MAINTENANCE SAFETY STANDARDS...
  • Page 5: General Information

    1 General Information 1.1 Overview Agilent’s high temperature sample preparation system is designed for routine dissolution and filtration of polymer solutions at temperatures up to 260 °C. Controlled heating and careful shaking of the samples assists dissolution, and a unique hand pipettor device, fitted with an interchangeable frit magazine, efficiently filters and dispenses the hot solution directly into autosampler vials with minimal handling.
  • Page 6: Unpacking

    1.3 Unpacking Care has been taken to ensure that the instrument should be received in proper condition. The packing and protection are designed for normal hazards of road, rail or air transit. Any damages to the container or instrument should be reported immediately to your local distributor or to Agilent, Inc.
  • Page 7: Utility Requirements

    Ensure that the power switch & appliance power connector situated at the rear of the unit  remain easily accessible at all times. Do not use this equipment in hazardous atmospheres or with hazardous materials for which it  was not designed. 1.5 Utility Requirements 1.5.1 Electrical Power...
  • Page 8 1.6.2 Preventative Maintenance Periodically inspect the power lead for signs of wear or damage. Contact Agilent’s Service Department to obtain a replacement lead – it must be fitted with a ferrite to maintain EMC compatibility. 1.6.3 Temperature The Sample Preparation System is designed to control the temperature of the sample preparation vials to temperatures up to 260 °C and therefore the metal hot plates, vials and the base of the hand pipettor will be extremely hot during normal operation.
  • Page 9: System Overview

    2 System Overview The Agilent 1260 Infinity II High Temperature Sample Prep System is illustrated in figures 2.1a and 2.1b. It comprises: A horizontal shaker [A]  Temperature controlled hot blocks [B] for sample dissolution in 20 mL vials. ...
  • Page 10 Figure 2.1b System Overview Agilent 1260 HT Sample Prep System Operation Manual...
  • Page 11 Displacement button to pump solution from one vial to another Finger lever to release frit magazine Frit Magazine containing 2 µm Stainless Steel Frits [J] Figure 2.2 Hand Pipettor Agilent 1260 HT Sample Prep System Operation Manual...
  • Page 12: Installing The Module

    3 Installing the module 3.1 Preparing the system 1. Place the system in the desired location. 2. Connect the power cable from the unit to the mains power supply. 3. Position the four hot blocks into the unit so that the large sample preparation vials are adjacent to the autosampler vials.
  • Page 13: System Operation

    4 System Operation 4.1 Power On The power switch is located at the rear of the instrument and when switched on power s the control system of the instrument. The power to the shaker and heaters are controlled by their respective on/off buttons on the control pad.
  • Page 14 4.2.3 Programming the Shaker Press Display Comments Toggle between mode sets using up or down arrow Shaker delay time displayed in minutes. Adjust using  and  keys Shaker time displayed in minutes. Adjust using  and  keys The above example programmes the shaker to shake for 20 minutes after an initial delay of 10 minutes.
  • Page 15 During the shaker cycle the shaker light will flash red if operating in ‘timed’ mode. In ‘continuous’ mode the shaker light remains red. Agilent 1260 HT Sample Prep System Operation Manual...
  • Page 16: Temperature Controller

    4.3 Temperature Controller 4.3.1 General The controller regulates the temperature of the heater which is positioned directly below the aluminium blocks containing the sample vials. The temperature is measured and fed back to the controller by means of a platinum resistance temperature sensor (RTD/PT100). The power fed to the heaters is then calculated by a standard PI algorithm, using control parameters which are accurately tuned.
  • Page 17: Safety Over-Temperature Cut-Out

    4.4 Safety Over-Temperature Cut-Out 4.4.1 General The temperature controller is protected against thermal run away but as a fail-safe for the unit an independent temperature measuring system is included that will cut the power to the heaters via a relay if the measured temperature exceeds the displayed over-temperature threshold. 4.4.2 Operation The overtemperature set point is displayed on the secondary display.
  • Page 18: Hand Pipettor

    4.4.3 Over-Temperature Cut-Out Test Procedure The over-temperature protection system should be tested regularly using the following procedure: 1. Turn the heaters OFF and allow the instrument to cool to below 50 °C 2. Adjust the cut-out temperature to 60 °C 3.
  • Page 19: Sample Preparation

    (TCB) for use in the Agilent Infinity II High Temperature GPC system operating at 160 °C, to illustrate the recommended use of the 1260 Infinity II High Temperature Sample Prep System. Adaptations of this procedure will be necessary to suit other samples or preferred procedures.
  • Page 20: Cleaning The Frits

    10. To use the next frit in the magazine, squeeze the hand lever to release the frit magazine and slide the magazine to the next position. Release the hand lever to seal the magazine in place. 11. The autosampler vials now need to be crimped and labelled in preparation to be transferred to the PL-GPC 220 carousel.
  • Page 21: Cleaning And Decontamination Procedure

    6 Cleaning and Decontamination Procedure Switch off and disconnect power cord from instrument before cleaning or decontamination. Allow any hot parts to cool before proceeding. Cleaning The exterior of the instrument should be cleaned by wiping down with a soft cloth moistened with dilute detergent solution, followed by wiping down with a cloth moistened with deionised water.
  • Page 22: Maintenance

    7 Maintenance Switch off and disconnect power cord from instrument before servicing. Allow any hot parts to cool before proceeding There are no user-serviceable parts inside the instrument. Agilent 1260 HT Sample Prep System Operation Manual...
  • Page 23 8 Appendix Safety Symbols Safety Symbols Table 10 Symbol Description The apparatus is marked with this symbol when the user should refer to the instruction manual in order to protect risk of harm to the operator and to protect the apparatus against damage. Indicates dangerous voltages.
  • Page 24: Safety Standards

    A WARNING WARNING alerts you to situations that could cause physical injury or death. ➔ Do not proceed beyond a warning until you have fully understood and met the indicated conditions. A CAUTION C A U T I O N alerts you to situations that could cause loss of data, or damage of equipment.
  • Page 25: Abstract

    To return unwanted products, contact your local Agilent office, or see www.agilent.com for more information. Radio Interference Never use cables other than the ones supplied by Agilent Technologies to ensure proper functionality and compliance with safety or EMC regulations. Test and Measurement...
  • Page 26 Agilent Technologies on Internet For the latest information on products and services visit our worldwide web site on the Internet at: http://www.agilent.com Agilent 1260 HT Sample Prep System Operation Manual...
  • Page 27 Infinity II High Temperature Sample Preparation System (G7823A). The manual describes the following: • General information, • Introduction • Site requirements • Unpacking • Using the module • Troubleshooting and diagnostics Agilent Technologies 2017 Printed in Germany 07/201 7 *G7823-91000* G7823-91000 Agilent Technologies...