Drive Timeout; Overcurrent Of Pump Drive - Agilent Technologies 1260 Infinity II User Manual

Flexible, infinitylab lc series
Hide thumbs Also See for 1260 Infinity II:
Table of Contents

Advertisement

Drive timeout

Error ID: 29204
Movement of drive during initialization is blocked mechanically.
• Parameter: 1 – 2 referring to pump drive
Probable cause
1
2

Overcurrent of pump drive

Error ID: 29202
The current consumption of the pump drive is too high.
• Parameter: 1 – 2 referring to pump drive
Probable cause
1
2
1260 Infinity II Flexible Pump User Manual
Blockage in flow path
Blockage of pump drive Drive motor defect.
Blockage of system before pressure sensor.
Drive motor defect.
Error Information
Pump Error Messages
Suggested actions
Remove capillary connection to system, check
outlet filter, check valves, check pump head.
Please contact your Agilent service
representative.
Suggested actions
Check for blockage of e.g. outlet valve filter frit,
Multi Purpose Valve, heat exchanger.
Please contact your Agilent service
representative.