Agilent Technologies 1260 Infinity II User Manual page 32

Flexible, infinitylab lc series
Hide thumbs Also See for 1260 Infinity II:
Table of Contents

Advertisement

2
Site Requirements and Specifications
Performance Specifications
Table 2
Feature
Instrument Control
Communications
Safety features and
maintenance
GLP features
Housing
32
Agilent 1260 Infinity II Flexible Pump (G7104C) Performance Specifications
Specification
LC & CE Drivers A.02.17 or above
Instrument Control Framework
(ICF) A.02.05 or above
InfinityLab LC Companion
(G7108AA) with firmware D.07.xx
or above
Instant Pilot (G4208A) with
firmware B.02.22 or above
Lab Advisor software B.02.10 or
above
Controller-area network (CAN),
RS232C, APG remote: ready, start,
stop, and shutdown signals, LAN
Leak detection, safe leak handling,
leak output signal for shutdown of
the pumping system.
No hazardous voltages in major
maintenance areas.
Extensive diagnostics, error
detection and display with Agilent
Lab Advisor software.
Early maintenance feedback (EMF)
for continuous tracking of
instrument usage in terms of seal
wear and volume of pumped mobile
phase with pre-defined and user
settable limits and feedback
messages.
Electronic records of maintenance
and errors.
All materials are recyclable.
Comments
For details about supported
software versions refer to the
compatibility matrix of your version
of the LC & CE Drivers
1260 Infinity II Flexible Pump User Manual