Timeout: Wait for Flow

Error ID: 29183
A target condition (flow rate) has been sent to the instrument which should
have been reached within an expected time frame but didn't. Either the limit,
time frame or the current value of the variable has been modified later directly
or indirectly.
Probable cause
1

Timeout: Wait for Pressure

Error ID: 29185
A target condition (pressure) has been sent to the instrument which should
have been reached within an expected time frame but didn't. Either the limit,
time frame or the current value of the variable has been modified later directly
or indirectly.
Probable cause
1
2
1260 Infinity II Flexible Pump User Manual
Incorrect parameters have been sent to the
instrument by the control software or
manual changes.
Incorrect parameters have been sent to the
instrument by the control software or
manual changes.
Leak
Error Information
Pump Error Messages
Suggested actions
Verify control software, macros, manual
commands.
Suggested actions
Verify control software, macros, manual
commands.
Run system pressure test for identifying and
localizing the leak. Tighten leak.