Agilent Technologies 1260 Infinity II User Manual page 163

Flexible, infinitylab lc series
Hide thumbs Also See for 1260 Infinity II:
Table of Contents

Advertisement

3
Remove the heat exchanger capillary by pushing the
connector up and pulling it out of the valve.
N O T E
A gold seal between outlet valve and heat exchanger
capillary is used for a tight connection.
5
Mount the pump head to the module. Do not fix the
screws at this stage!
1260 Infinity II Flexible Pump User Manual
4
Unscrew the outlet valve with a 14 mm wrench (1) and
remove it (2).
6
Insert the new outlet valve and tighten it using a torque
wrench with a 14 mm bit set to 10 Nm.
Maintenance
Replace the Outlet Valve
7
163