7
Maintenance

Remove the Jet Weaver

Remove the Jet Weaver

Tools required
p/n
8710-0510
8710-0899
5023-2502
Parts required
p/n
0100-1259
G4204-04002
Preparations
• Select Do not use mixer in ChemStation.
• Switch off the pump at the main power switch.
1
Remove capillary connections from the Jet Weaver to
the Multi Purpose Valve.
134
Description
Wrench open 1/4 — 5/16 inch
Pozidriv screwdriver
Hex driver SW-6.35, slitted
Description
Plastic fittings
Cover RFID Tag
Metal lid for Jet Weaver
2
Plug the valve ports 1 and 2 with two plastic fittings.