Pump Drive Stop Not Found; Timeout: Wait For Composition - Agilent Technologies 1260 Infinity II User Manual

Flexible, infinitylab lc series
Hide thumbs Also See for 1260 Infinity II:
Table of Contents

Advertisement

Pump drive stop not found

Error ID: 29207
The pump drive stop has not been found.
• Parameter: 1 – 2 referring to pump drive
Probable cause
1

Timeout: Wait for Composition

Error ID: 29180
A target condition (composition) has been sent to the instrument which
should have been reached within an expected time frame but didn't. Either the
limit, time frame or the current value of the variable has been modified later
directly or indirectly.
Probable cause
1
1260 Infinity II Flexible Pump User Manual
Pump drive spindle is defect.
Incorrect parameters have been sent to the
instrument by the control software or
manual changes.
Error Information
Pump Error Messages
Suggested actions
Please contact your Agilent service
representative.
Suggested actions
Verify control software, macros, manual
commands.
6
113

Advertisement

Table of Contents
loading

Table of Contents