11
Place the seal holder onto the spring housing.
N O T E
Mind the correct orientation: The backup ring must
face upwards and the seal holder must sit correcly.
13
Place the screws into the pump heads and loosely
tighten them in a crosswise manner.
N O T E
The spring housing will tilt slightly when the first
screw is hand tightened. Stop at this point and
continue to tighten the three other screws in a
crosswise manner.
1260 Infinity II Flexible Pump User Manual
Pump Head Maintenance (Tool Free)
12
Place the assembled spring housings on top of the pump
chambers.
N O T E
Both spring housings are identical, there is no risk
when mixing them, but make sure that the seal holder
is oriented correctly.
14
Mind the correct orientation of the link plate and click it
into place.
N O T E
The Pump Head Holder has a marker to illustrate the
correct placement of the link plate. The link plate
holds an identification tag; this has to be placed onto
the correct position to be readable by the pump.
7
Maintenance
157