Using the Module
Setting up the Pump with the Instrument Control Interface
Instrument Signals
The instrument signals are displayed when clicking on the icon in the upper
right corner of the module dashboard. The following pump signals are
displayed:
Flow
Pressure
Tuning
Pressure Limit
Composition A:B
Composition C:D
Valve position
Prepare Pump
60
The current solvent flow rate (in mL/min).
The current pump ressure (in bar, psi or MPa, see
"Instrument Configuration"
Indicates the tuning efforts of the pump. For pumps operating
as expected, the signal should stay in a range of -1 to +1
within the full scale of -2 to +2.
The current maximum pressure limit.
The contributions of channels A and B to the current solvent
composition.
The contributions of channels C and D to the current solvent
composition.
The current valve position.