Special Settings

The special settings are required for specific actions (normally in a service case).
The tables include both settings for modules – with on-board LAN and without on-board
N O T E
LAN. They are identified as LAN and no LAN.
Boot-Resident
Firmware update procedures may require this mode in case of firmware
loading errors (main firmware part).
If you use the following switch settings and power the instrument up again, the
instrument firmware stays in the resident mode. It is not operable as a module. It
only uses basic functions of the operating system for example, for communication.
In this mode the main firmware can be loaded (using update utilities).
Table 14
Boot Resident Settings (On-board LAN)
Mode Select
TEST/BOOT
Forced Cold Start
A forced cold start can be used to bring the module into a defined mode with
default parameter settings.
Loss of data
C AU T I O N
Forced cold start erases all methods and data stored in the non-volatile memory.
Exceptions are calibration settings, diagnosis and repair log books which will not be
erased.
If you use the following switch settings and power the instrument up again, a
forced cold start has been completed.
Table 15
Forced Cold Start Settings (On-board LAN)
Mode Select
TEST/BOOT
1260 Infinity II Flexible Pump User Manual
SW1
SW2
SW3
1
1
Save your methods and data before executing a forced cold start.
SW1
SW2
SW3
1
1
Setting the 8-bit Configuration Switch
SW4
SW5
1
0
0
SW4
SW5
0
0
0
Hardware Information
SW6
SW7
SW8
0
0
SW6
SW7
SW8
0
0
10
0
1
221