3
Unscrew the high pressure outlet filter.
5
Reassemble the high pressure outlet filter (approx.
16 Nm).
7
Mount the capillary connection to the pressure sensor
(1) and to the Multipurpose valve (2).
1260 Infinity II Flexible Pump User Manual
Replace the High Pressure Outlet Filter or Filter Frit
4
Replace the filter frit.
N O T E
Remove the old filter frit with the sharp side of the
blade of a knife.
6
Reinstall the high pressure outlet filter to the filter
holder.
7
Maintenance
169