• Outlet valve not properly assembled
Dynamic Leak > 30 µL/min or Dynamic Leak Rate Test fails
• Air in the primary pump chamber
• Abort due to over pressure
• Leak in Inlet Valve
• Outlet valve not properly assembled
• Leaky piston seals and/or position dependent leaks on the piston
1260 Infinity II Flexible Pump User Manual
Remove waste lines from the automatic purge valve
Damage to rotor seal and/or stator head
Re-tighten the outlet valve
Check the position of the gold seal
Check for air in the solvent inlet lines and the Tuning signal
Purge the lines, Prime and Condition the pump head
Check solvent and solvent settings
Purge and condition the pump head with water
Check for moving air bubbles in tubing directly to the Inlet Valve
Purge the lines with water to remove dirt
Knock at the valve, clean or replace it
Re-tighten the outlet valve
Check the position of the gold seal
Remove the SW tubes from the support ring and check for leaks
Replace the piston seals and clean the pistons
Ensure that seals are lubricated when pushed in
Use abrasive mesh >5000 grit
Troubleshooting and Diagnostics
Pump Leak Rate Test