Maintenance
Replace the Pressure Sensor
3
Carefully pull out the pressure sensor for about 2 cm.
Then unscrew the cable from the pressure sensor.
5
Fix the pressure sensor to the instrument chassis.
130
4
Connect the new pressure sensor to the pressure sensor
connector.
6
Connect the capillary from the pump head outlet to the
pressure sensor inlet (1). Connect the capillary from the
high pressure filter to the pressure sensor outlet (2). Two
arrow signs on the pressure sensor indicate the flow
direction.
1260 Infinity II Flexible Pump User Manual