Replace The High Pressure Outlet Filter Or Filter Frit - Agilent Technologies 1260 Infinity II User Manual

Flexible, infinitylab lc series
Hide thumbs Also See for 1260 Infinity II:
Table of Contents

Advertisement

7
Maintenance

Replace the High Pressure Outlet Filter or Filter Frit

Replace the High Pressure Outlet Filter or Filter Frit

When
For removing blockages and leaks in the high pressure filter assembly. The outlet filter should be
replaced as required depending on the system usage.
Tools required
p/n
5023-2502
8710-0510
8710-1924
5067-5688
5067-5690
Parts required
#
1
OR
1
1
Remove the capillary from the high pressure outlet filter
to the pressure sensor (1) and from the high pressure
outlet filter to the Multipurpose valve (2).
168
Description
Hex driver SW-6.35, slitted
Wrench open 1/4 — 5/16 inch
Wrench open 14 mm
Torque wrench 1 – 25 Nm with 14 mm wrench
Torque wrench head, 14 mm, for torque wrench
p/n
Description
G4204-60004
Outlet filter Quaternary Pump/Flexible Pump
5067-5716
Frit for pump outlet filter 2/pk
2
Remove the high pressure outlet filter from the filter
holder.