Agilent Technologies 1260 Infinity II User Manual page 43

Flexible, infinitylab lc series
Hide thumbs Also See for 1260 Infinity II:
Table of Contents

Advertisement

3
Using the Module
Flushing the Filter
In the instrument control panel of Agilent user interfaces, use the context
menu and select Flush Filter On, see also
"Context Menu"
on page 61.
Optional inline filter (for G7104A/C, G4204A)
The pump can be equipped with an additional inline filter ( Inline Filter
Assembly (5067-5407)) with a nominal filter pore size of 0.3 μm.
Advantages of the inline filter:
• Very small internal volume
• Specified for working at high pressures
• Possibility of back- flushing the filter
Using the inline filter is recommended:
• To protect the downstream system from blockages,
• For solvent combinations that can form precipitation after mixing,
• For applications running with buffers.
General hints for effective usage of the inline filter:
• Filter solvents before usage,
• Follow best practices,
• Back- flush the filter weekly, (Flush Filter in context menu figure)
• Exchange the filter frit on a regular basis. ( Frit 0.3 μm for inline filter, 5/pk
(5023-0271))
See Technote G4226-90131 for further reference.
N O T E
1260 Infinity II Flexible Pump User Manual