Writing The Pump Encoder Tag Failed; Pump Drive Blocked Or Encoder Failed - Agilent Technologies 1260 Infinity II User Manual

Flexible, infinitylab lc series
Hide thumbs Also See for 1260 Infinity II:
Table of Contents

Advertisement

Writing the pump encoder tag failed

Error ID: 29200
Writing the pump encoder tag has failed.
• Parameter: 1 – 2 referring to pump drive
Probable cause
1
2

Pump drive blocked or encoder failed

Error ID: 29214
Pump drive blocked or encoder failed.
• Parameter: None
Probable cause
1
1260 Infinity II Flexible Pump User Manual
Defect connection between encoder and
main board.
Defect tag Defect connection between tag
and encoder.
Blockage of the pump drive Drive encoder
failed.
Error Information
Pump Error Messages
Suggested actions
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Suggested actions
Please contact your Agilent service
representative.
6
105

Advertisement

Table of Contents
loading

Table of Contents