Maintenance
Pump Head Maintenance (Tool Free)
7
Take care that the Seal Handling Device is seating flush
and press the wash seal into the seal holder.
N O T E
The Seal Handling Device has a cavity to fit over the
pins of the seal wash tubings.
9
Remove the seal holders from the pump chambers.
156
8
Repeat for the other seal holder.
10
Lubricate the piston with isopropanol and place it into
the spring housing.