3
Insert the Jet Weaver to the opening in the front panel
(1) and push it down (2).
5
Mount both capillary connections to the Jet Weaver
observing the correct orientation (long fitting to Jet
Weaver).
7
Configure the Jet Weaver as mixer in the user interface.
1260 Infinity II Flexible Pump User Manual
4
Fix the Jet Weaver with a screw.
6
Connect the inlet capillary of the Jet Weaver to port 2 of
the Multi Purpose Valve. Connect the outlet capillary to
port 1.
Maintenance
Install the Jet Weaver
7