Replace the Inlet Weaver

Parts required
p/n
G4204-81090
Preparations
• Switch off pump at the main power switch
• Open the doors
• Use an optional solvent shutoff valve or lift up solvent filters inside solvent bottles for avoiding
• For easy access to the inlet weaver assembly, remove tubing connections between MCGV and
1
Open the screw at the bottom of the inlet valve.
3
Pull the Inlet Weaver out of the Inlet Valve.
1260 Infinity II Flexible Pump User Manual
Description
Quaternary Pump/Flexible Pump Inlet Weaver Assembly
leakages
degasser
2
Open the fitting at the center of the multi-channel
gradient valve (MCGV). Remove the inlet weaver from
the MCGV.
4
Insert the new inlet weaver to the inlet valve. Fix the
weaver with the plastic screw.
Maintenance

Replace the Inlet Weaver

7
131