6
Error Information
Pump Error Messages
Drive Encoder failed
Error ID: 29210
Drive encoder failed during pump drive calibration.
Probable cause
1
Drive phases differ too much in electric resistance
Error ID: 29216
Pump drive calibration has failed due to a strong difference electric
resistances for different motor phases.
Probable cause
1
2
3
Degasser's pressure limit violation
Error ID: 29220
Pressure too far above the limit.
Probable cause
1
2
3
116
Internal error.
Pump drive cable defect.
Pump drive defect.
Defective main board.
Leak in degasser chamber or degasser
tubing.
Defect vacuum pump.
Degasser chamber empty or connected to air. Block unused degasser channels.
Suggested actions
Contact Agilent support.
Suggested actions
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Suggested actions
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
1260 Infinity II Flexible Pump User Manual